A study for high accuracy real-time 3D ultrasonic location system

Research output: Contribution to journalArticlepeer-review

1 Scopus citations

Abstract

We discussed a high accuracy real-time 3D ultrasonic location system in this article. The signal received was sampled after it passed the TGC and the logarithmic amplifier. Inside the DSP, we used the dynamic threshold tracing technique to improve the accuracy. The result was processed with Weighted Arithmetic Average. By testing the 40 kHz 3D location system, we have arrived at the accuracy of 1 cm.

Original languageEnglish
Pages (from-to)e1499-e1502
JournalUltrasonics
Volume44
Issue numberSUPPL.
DOIs
StatePublished - 22 Dec 2006
Externally publishedYes

Keywords

  • Beam angle
  • Doppler
  • Dynamic threshold tracing
  • Real-time
  • TGC

Fingerprint

Dive into the research topics of 'A study for high accuracy real-time 3D ultrasonic location system'. Together they form a unique fingerprint.

Cite this