N2 - This paper presents a noncontact capacitive sensing system (C-Sens) for locomotion mode recognition of transtibial amputees. C-Sens detects changes in physical distance between the residual limb and the prosthesis. The sensing front ends are built into the prosthetic socket without contacting the skin. This novel signal source improves the usability of locomotion mode recognition systems based on electromyography (EMG) signals and systems based on capacitance signals obtained from skin contact. To evaluate the performance of C-Sens, we carried out experiments among six transtibial amputees with varying levels of amputation when they engaged in six common locomotive activities. The capacitance signals were consistent and stereotypical for different locomotion modes. Importantly, we were able to obtain sufficiently informative signals even for amputees with severe muscle atrophy (i.e., amputees lacking of quality EMG from shank muscles for mode classification). With phase-dependent quadratic classifier and selected feature set, the proposed system was capable of making continuous judgments about locomotion modes with an average accuracy of 96.3% and 94.8% for swing phase and stance phase, respectively (Experiment 1). Furthermore, the system was able to achieve satisfactory recognition performance after the subjects redonned the socket (Experiment 2). We also validated that C-Sens was robust to load bearing changes when amputees carried 5-kg weights during activities (Experiment 3). These results suggest that noncontact capacitive sensing is capable of circumventing practical problems of EMG systems without sacrificing performance and it is, thus, promising for automatic recognition of human motion intent for controlling powered prostheses.
AB - This paper presents a noncontact capacitive sensing system (C-Sens) for locomotion mode recognition of transtibial amputees. C-Sens detects changes in physical distance between the residual limb and the prosthesis. The sensing front ends are built into the prosthetic socket without contacting the skin. This novel signal source improves the usability of locomotion mode recognition systems based on electromyography (EMG) signals and systems based on capacitance signals obtained from skin contact. To evaluate the performance of C-Sens, we carried out experiments among six transtibial amputees with varying levels of amputation when they engaged in six common locomotive activities. The capacitance signals were consistent and stereotypical for different locomotion modes. Importantly, we were able to obtain sufficiently informative signals even for amputees with severe muscle atrophy (i.e., amputees lacking of quality EMG from shank muscles for mode classification). With phase-dependent quadratic classifier and selected feature set, the proposed system was capable of making continuous judgments about locomotion modes with an average accuracy of 96.3% and 94.8% for swing phase and stance phase, respectively (Experiment 1). Furthermore, the system was able to achieve satisfactory recognition performance after the subjects redonned the socket (Experiment 2). We also validated that C-Sens was robust to load bearing changes when amputees carried 5-kg weights during activities (Experiment 3). These results suggest that noncontact capacitive sensing is capable of circumventing practical problems of EMG systems without sacrificing performance and it is, thus, promising for automatic recognition of human motion intent for controlling powered prostheses.
KW - Capacitive sensing
KW - locomotionmode recognition
KW - lower-limb prosthesis
KW - noncontact wearable sensing
KW - quadratic discriminant classifier
UR - https://www.scopus.com/pages/publications/84912124595
U2 - 10.1109/TBME.2014.2334316
DO - 10.1109/TBME.2014.2334316
M3 - 文章
C2 - 25014949
AN - SCOPUS:84912124595
SN - 0018-9294
VL - 61
SP - 2911
EP - 2920
JO - IEEE Transactions on Biomedical Engineering
JF - IEEE Transactions on Biomedical Engineering
IS - 12
M1 - 6847122
ER -