Eliminating cardiac contamination from myoelectric control signals developed by targeted muscle reinnervation

  • Rehabilitation Institute of Chicago
  • Northwestern University Feinberg School of Medicine
  • Northwestern University

科研成果: 期刊稿件文章同行评审

49 引用 (Scopus)

摘要

The electrocardiogram (ECG) artifact is a major noise contaminating the myoelectric control signals when using shoulder disarticulation prosthesis. This is an even more significant problem with targeted muscle reinnervation to develop additional myoelectric sites for improved prosthesis control in a bilateral amputee at shoulder disarticulation level. This study aims at removal of ECG artifacts from the myoelectric prosthesis control signals produced from targeted muscle reinnervation. Three ECG artifact removal methods based on template subtracting, wavelet thresholding and adaptive filtering were investigated, respectively. Surface EMG signals were recorded from the reinnervated pectoralis muscles of the amputee. As a key parameter for clinical myoelectric prosthesis control, the amplitude measurement of the signal was used as a performance indicator to evaluate the proposed methods. The feasibility of the different methods for clinical application was also investigated with consideration of the clinical speed requirements and memory limitations of commercial prosthesis controllers.